Загрузил Тимофей Августинович

Гехалтеметр-OGM-EN-V01.05-01

реклама
HAFFMANS
INSTRUCTION MANUAL
IN-LINE O2 GEHALTEMETER
TYPE OGM
W W W.HAFFMANS.NL
OGM serial number: from Control Unit OGM-11001
OGM-31001
P30101
P50501
OGM from software version: V3.01-01
Read this instruction manual carefully
before using the instrument.
Haffmans BV is not responsible for any damage caused by the owner’s failure to follow the instructions given in this
manual. Haffmans BV is not responsible for print or translation errors in this edition of the manual. The specifications
and design of this product are subject of change without notice. All rights reserved. No part of this publication may be
reproduced or distributed in any form or by any means, or stored in a database or retrieval system, without the prior
written permission of Haffmans BV.
Art. No.
Release date
Version
:
:
:
112.364
01 November 2014
V01.05-01
HAFFMANS OGM
Contents
1.
Technical specifications ............................................................................................... 7
2.
Introduction ................................................................................................................. 11
2.1
Intended use ..................................................................................................... 11
3.
Safety ........................................................................................................................... 13
3.1
Symbols ........................................................................................................... 13
3.2
Precautionary measures and safety instructions.......................................... 13
4.
General product information ...................................................................................... 15
4.1
Control unit ....................................................................................................... 15
4.2
Sensor .............................................................................................................. 15
4.3
Coating holder ................................................................................................. 16
4.4
Operation ......................................................................................................... 16
4.5
O2 -measurement ............................................................................................. 17
5.
Installation ................................................................................................................... 19
5.1
Contents of the delivery .................................................................................. 19
5.2
Assembly of the OGM ...................................................................................... 20
5.3
Preparation and starting up ............................................................................ 21
5.4
Main measuring screen ................................................................................... 21
5.5
Using for the first time .................................................................................... 22
6.
Operation ..................................................................................................................... 31
6.1
Operation OGM ................................................................................................ 31
6.2
Local and Remote control ............................................................................... 33
6.3
Logbook ........................................................................................................... 33
7.
Calibration ................................................................................................................... 35
7.1
Calibration of the temperature sensor ........................................................... 35
7.2
Calibration of the O2 sensor(s) ....................................................................... 35
7.3
Calibrating analogue outputs ......................................................................... 41
8.
Cleaning and maintenance ......................................................................................... 43
8.1
Daily maintenance ........................................................................................... 43
9.
Faults and repair ......................................................................................................... 45
9.1
List of faults ..................................................................................................... 45
9.2
Alarm messages .............................................................................................. 45
9.3
Repairs and maintenance ............................................................................... 47
10.
Deactivating, storage and transport .......................................................................... 49
10.1 Deactivating ..................................................................................................... 49
10.2 Storage ............................................................................................................. 49
10.3 Transport .......................................................................................................... 49
OGM
Page 5 of 81
HAFFMANS OGM
Appendix A: Formulas .......................................................................................................... 51
Appendix B: Software structure ........................................................................................... 53
Appendix C: Connections ..................................................................................................... 57
Appendix D: RS-protocol ...................................................................................................... 63
Appendix E: Instruction for assembly panel ....................................................................... 65
Appendix F: LHG Sensor (Option) ....................................................................................... 69
F.1
Drawing ............................................................................................................ 69
F.2
Mounting instructions ..................................................................................... 70
Appendix G: Customer service ............................................................................................ 71
Appendix H: Warranty conditions ........................................................................................ 73
Appendix I: EC Declaration of conformity .......................................................................... 75
Appendix J: Spare parts ....................................................................................................... 77
Appendix K: PROFIBUS ........................................................................................................ 81
Page 6 of 81
OGM
HAFFMANS OGM
1.
Technical specifications
Scope of
measurements
O2 LHO
O2 WLO
O2 LHG
Temperature
Accuracy*
: standard
: [ppb w/w]
: 0 tot 2000 (DO)
: available
: [µg/l]
: 0 tot 2000 (DO)
: [ppm w/w]
: 0.00 tot 2.00 (DO)
: [mg/l]
: 0.00 tot 2.00 (DO)
: [%] O2
: 0,000 to 4,180 (gas)
: [%a.s.]
: 0.00 to 20.0 (gas)
: [ppm
vol/vol]
: 0 to 41800 (gas)
: standard
: [ppb w/w]
: 0 tot 45000 (DO)
: available
: [mg/l]
: 0.00 tot 45.00 (DO)
: [µg/l]
: [ppm w/w]
: 0 tot 45000 (DO)
: 0.00 tot 45.00 (DO)
: [%] O2
: 0.00 to 100.00 (gas)
: [%a.s.]
: 0.00 to 500.00 (gas)
: [ppm
vol/vol]
: 0 to 1000000 (gas)
: standard
: [ppm
vol/vol]
: 0 to 200 (gas)
: available
: [%] O2
: 0 to 0.02 (gas)
: [% a.s.]
: 0 to 0.1 (gas)
: standard
: [oC]
: -5.0 to 40.0
: available
: [oF]
: 23.0 to 104.0
: [ppb]
: ± Approx. (1 ppb+2% of the
measured value)
: ± (1 µg/l +2% of the measured
value)
: LHO
: O2 dissolved
(DO)
: [µg/l]
OGM
: [ppm]
: ± (0.001 ppm+2% of the measured
value)
: [mg/l]
: ± (0.001 mg/l+2% of the measured
value)
Page 7 of 81
HAFFMANS OGM
O2 in gas
: [% a.s.]
: ± (0.01%a.s.+2% of the measured
value )
: [%] O2
: ± (0.002 %O2 +2% of the measured
value )
: [ppm]
: ± (20 ppm +2% of the measured
value )
: [mg/l]
: 0 – 45 (< 5% of the measured value
from 0.1 mg/l)
: [ppb]
: 0 – 45.000 (< 5% of the measured
value from 100 ppb)
: [µg/l]
: 0 – 45.000 (< 5% of the measured
value from 100 µg/l)
: [ppm]
: 0 – 45 (< 5% of the measured value
from 0.1 ppm)
: [% a.s.]
: 0 – 45 (< 5% of the measured value
from 1.098 ppm)
: [%] O2
: 0 – 45 (< 5% of the measured value
from 0.230 ppm)
: [ppm]
: 0 – 1,000,000 (< 5 % of the
measured value from 2300 ppm)
: [ppm]
: 0 – 200 (2 ppm.+/- 5% vdgw)
: [%] O2
: 0 – 0.02 (0.0002 % O2 +/- 5% vdgw)
: [% a.s.]
: 0 – 0.1 (0.001 %a.s +/- 5% vdgw)
: [°C]
: ± 0,1°C
: [°F]
: ± 0,2°F
: [barg]
: 10.00 barg
WLO
: O2 dissolved
(DO)
O2 in gas
LHG
: O2 in gas
: temperature
Maximum
pressure during
use
: [psi]
: 145.0 psi
2
Product speed
: [kg/cm ]
: 10.20 kg/cm2
: [kPa]
: 1000 kPa
: min
: [m/s]
: >0
: max
: [m/s]
:-
during measurement
In pipe
Page 8 of 81
OGM
HAFFMANS OGM
Maximum
cleaning
LHO/WLO
: [°C]
: 110
: [°F]
: 230
: [°C]
: 50
: [°F]
: 120
: max
: [m/s]
: 10m/s
: O2
: [s]
standard
: [s]
: Adjustable from each 2 to 999 s or
auto mode
: 30s
temperature
LHG
Speed of cleaning
liquid during CIP
Measuring
frequency
: 400 measurements / O2 sensor
Memory
Power
Control unit: : standard
: optional
: 100-240V ~ 50-60Hz, 25 W
: 24 V DC, 25 W
Installation
O2 sensor : LHO/WLO
: Varivent (>DN40) of 25 mm
: LHG
: 25 mm
Control unit: : standard
: wall assembly mounted
: optional
: panel assembly or ,
pipe/ sensor assembly mounted
IP class
O2 sensor : standard
: IP 67
: optional
: IP 68
: IP 65
Control unit:
Dimensions
O2 sensor : Varivent
: 25mm
Control unit:
: mm
: Ø 84x175 excl. plug
: mm
: Ø 84 x 240 excl. plug
: mm
: L235 x W205 x H165
: Kg
: 1.8 kg
: Kg
: 2.2 kg
: Kg
: 2.0 kg
Weight
O2 sensor : Varivent
: 25mm
Control unit:
OGM
Page 9 of 81
HAFFMANS OGM
Max. number of
O2 sensors for
each control unit
:2
* The accuracy from the OGM is determined at 20°C
Page 10 of 81
OGM
HAFFMANS OGM
2.
Introduction
2.1
Intended use
The OGM (O2 content gauge) is an in-line oxygen content gauge for measuring the oxygen
content in gas or the dissolved oxygen content in liquids. The standard gauge consists of a
control box and a sensor. A maximum of two sensors can be connected to each control box. A
second sensor is available as an optional extra.
The OGM must be installed in a product pipe. The connections can be: Varivent/viewing glass
from DN40 or a 25 mm welding socket. The dissolved oxygen must be measured in a full,
flowing product pipe. Using the sensor in a pipe that is not full or taking the measurement whilst
the product is not flowing will not cause errors nor will it cause damage to the device.
The use of the Varivent connection will be assumed throughout this instruction manual. If an
explanation or a part of the text specifically refers to the 25 mm connection, this will be indicated
by the use of an asterisk (*).
OGM
Page 11 of 81
HAFFMANS OGM
Page 12 of 81
OGM
HAFFMANS OGM
3.
Safety
3.1
Symbols
The symbols ‘Note!’, ‘Warning’ and ‘Danger’ used in this instruction manual have the following
meanings.
Note!
Instructions for the correct and effective use of the instrument.
Warning
Incorrect or careless use may cause serious damage to the
instrument.
Incorrect or careless use may place the user or the
surroundings in DANGER.
3.2
Precautionary measures and safety instructions
Warning
The use of chlorite-based or fluorine-based cleaning products
is not permitted. These may cause damage to the instrument.
Ensure that the pipe is not pressurized and that the instrument
is disconnected from the power supply during assembly and
disassembly. This will prevent electric shocks.
During CIP/cleaning, the outside of the instrument may
become hot. The sensor will turn itself off automatically when
the inside of the sensor reaches a temperature of 65°° C. Do not
touch the outside of the instrument!
OGM
Page 13 of 81
HAFFMANS OGM
Page 14 of 81
OGM
HAFFMANS OGM
4.
General product information
4.1
Control unit
“ENTER” button
“CURSOR UP”/”SHORTCUT” button
“Start” button
“CURSOR DOWN”/”Sensor” button
Fig 4.1: Operating panel
4.2
Sensor
Coating holder
Temperature sensor
Fig. 4.2: Sensor varivent connection
OGM
Fig. 4.3: Sensor 25mm
Page 15 of 81
HAFFMANS OGM
4.3
Coating holder
The OGM Varivent and the 25 mm connection can both be delivered in the versions LHO and
WLO.
The LHO coating has a six-sided head. The WLO version has a square head. The LHG can be
recognized by the six-sided head and the metal grid on the coating holder. See following
images.
Fig. 4.4: LHO coatingholder
Fig: 4.5: WLO coatingholder
Fig: 4.6: LHG coatingholder
The type is also indicated on the coating holder.
4.4
Operation
The instrument can be operated:
a. Locally with the control unit by the buttons on the panel
b. Remote with I/O or optional with Profibus
For further operational instructions, please see chapter 6
Page 16 of 81
OGM
HAFFMANS OGM
4.5
O2 -measurement
The O2 measurement is based on measuring the luminescence of an oxygen-sensitive layer.
The luminescence changes according to the oxygen partial pressure. The amount of dissolved
oxygen gas in the liquid is calculated, using the measured oxygen partial pressure and
temperature.
The oxygen sensor optically measures the O2 level of the liquid, using the luminescence
measuring principle, whereby an oxygen-sensitive layer is exposed to blue light, and, as a
result, molecules in the oxygen-sensitive layer are excited.
Fig. 4.7
1) Luminescence in the absence of oxygen
2) Luminescence in the presence of oxygen
In the absence of oxygen, the molecules light up red. If oxygen is present, then the oxygen
molecules collide with the molecules in the oxygen-sensitive layer. The molecules that collide
with oxygen no longer light up. As a result of this process, a relationship is created between the
oxygen concentration and both the light intensity and the speed at which the light intensity is
reduced. The light intensity reduces at a higher oxygen concentration and the speed at which
the intensity reduces increases as the concentration increases.
The oxygen content is calculated using the time difference between the exposure to the blue
light and the molecules lighting up (phase shift) and the product temperature.
OGM
Page 17 of 81
HAFFMANS OGM
Page 18 of 81
OGM
HAFFMANS OGM
5.
Installation
5.1
Contents of the delivery
The OGM has been calibrated, checked and tested by Haffmans BV before it was shipped, and
the instrument is therefore ready for immediate use. You do not have to calibrate the instrument
again before starting operations.
Check whether the delivery is complete and undamaged. If the delivery is incomplete or
damaged, please contact Haffmans BV or the Haffmans agency in your area immediately (also
see: www.Haffmans.nl). In all correspondence, please note the OGM’s serial number, order
number or invoice number (as provided by Haffmans BV).
Note!
Before using the instrument, please insure that the instrument is
complete and no parts are missing.
The delivery includes:
•
•
•
•
•
•
•
•
•
Control box
O2 sensor
Power cable
Sensor communication cable
I/O cable for analogue outputs
Manual + certification calibration data sensor
Calibration beaker, including additional o-ring 60x3 (* 20x2,5)
Attachment clip for wall mounting
RS-232 Cable
You can also order, optionally:
•
•
•
•
Sensor / Pipe mounting option.
Retaining ring for connecting the sensor for the Varivent assembly
CD-Rom with PC software to read out data, incl. 3m RS cable
Inline pipe section with 2 varivent connections, one with inspection glass and
one with retaining ring + o-ring. The following sizes are available DN40, DN50,
DN65, DN80, DN100, DN125, DN150.
For other spare parts and components see Appendix J.
OGM
Page 19 of 81
HAFFMANS OGM
5.2
Assembly of the OGM
5.2.1 Assembly control box
There are three options to install te control box.
1. Wall mounting
2. Panel mounting
3. Pipe mounting
For more information and instructions see appendix E.
5.2.2 Assembly sensor
Assembly advice horizontal piping
Always build in Varivent + 25 mm sensors horizontally.
V
X
Fig. 5.2: Assembly advice sensors
Assembly advice vertical piping
Rotation / position of sensors irrelevant.
Note!
Page 20 of 81
Bends and stream impact do not have to be taken into account.
Ensure at all times that (irrespective of the position of the
sensor) the coating is fully in the product stream. If this condition
is not complied with, you will always measure a faulty oxygen
value.
OGM
HAFFMANS OGM
5.3
Preparation and starting up
Depending on the version of the OGM ordered, this shall be connected as shown in Appendix
C. After all cables have been connected, the power can be turned on and the OGM starts up
immediately.
Warning
Hard materials, such as woodchips and/or grindings can
damage the OGM’s measuring head.
Warning
First carry out a cleaning cycle after the instrument has been
turned on.
5.4
Main measuring screen
Three seconds after you have turned on the power, the main measuring screen appears.
Depending on the number of sensors connected, the following measuring screens
will appear (after the control box has received the calibration data, shown by the INIT message
on the bottom of the display):
Fig. 5.4: One sensor connected
Fig 5.5: Two sensors connected
In advance, you have to set in the menu configuration sensor settings, how many sensors
are connected and if these apply to gas or liquid, see chapter 5.5.2. “sensor settings”. The
control unit then has to be restarted. Sensor 1 is set standard to ON and sensor 2 to OFF.
When you start up the instrument for the first time, the main measuring screen shall appear, see
fig. 5.3.
When two sensors are connected/turned on, the measuring screen for two sensors (see fig. 5.4)
appears as main screen, and you can scroll with the ↓ button to the measuring screen of sensor
1 or sensor 2. If one sensor is turned on, only the measuring screen of that sensor appears and
you cannot scroll down. If two sensors are connected and one or both do not give measurment
values, then a dotted line will be displayed and the status notification changes (see fig. 5.3).
The top line states the real time and the control mode (remote/local). In the bottom of the screen
a line – XXXXXXXX – is activated, which can show an alarm/message, if applicable. If there are
OGM
Page 21 of 81
HAFFMANS OGM
multiple alarms, these will scroll into this line consecutively for 3 seconds. The bottom line
shows the status of the O2 measurements.
Possible modes:
INIT/MEASURE/STANDBY/CALIBRATION
Immediately after starting up the instrument, INIT appears at the bottom of the screen. This
means that the control unit is loading the calibration values of the sensors. This will take approx.
30 seconds after the sensor has been switched on. When in standby mode, no O2
measurements are conducted and lines appear at the O2 value. The temperature of that sensor
is indicated permanently. The calibration mode is activated when the measuring cell is
calibrated. You cannot measure when this is ongoing.
To receive a quick information update, go to the SHORTCUT menu with the ↑ button, to view
additional measuring values, charts and alarms.
Enter-button:
If you press the enter-button for 3 seconds in the measuring screen, you will access the main
menu screen.
5.5
Using for the first time
Before starting operations with the OGM, the manufacturer’s default setting can be changed to
your preferential settings. The OGM has two operator levels:
- Operator level
(A)
- Supervisor level
(B)
Appendix B “Software Structure” describes the access of both levels to the various screens.
5.5.1 User level
To access the supervisor level, please follow the following steps.
1. Push the “Enter” button in the main measuring screen and hold this for three seconds.
The main menu appears.
2. Select “User” with the cursor buttons and confirm the selection by pressing “Enter”.
The password menu appears.
3. Select “Password” with the cursor buttons and confirm the selection by pressing “Enter”.
4. Enter the right password with the cursor buttons in [6848] and confirm this with “Enter”.
5. A message with “Correct Input” or “Incorrect Input” appears after the code has been
entered and confirmed.
6. Then select the
icon with the cursor buttons and confirm with “Enter”.
7. All screens in the Operator and Supervisor level are now visible.
8. If you forget to log off, or if no buttons are used during 30 minutes in the Supervisor
level, the software automatically switches back to the Operator level. This time cannot
be adjusted.
Page 22 of 81
OGM
HAFFMANS OGM
5.5.2 Configuration
If wanted, the configuration of the OGM can be adjusted.
To access the configuration menu, please follow the following steps.
1. Push the “Enter” button in the main measuring screen and hold this for three seconds.
The main menu appears.
2. Select “Configuration” with the cursor buttons and confirm the selection by pressing
“Enter”.
The configuration menu appears.
3. The following settings can be adjusted in this menu.
Units
The units used in the OGM for the O2 values and temperature can be selected optionally.
1. Select the submenu “Units” in the configuration menu with the cursor buttons and
confirm with “Enter”.
2. Then select sensor 1 or 2 or the temperature and confirm with “Enter”.
3. With the arrow buttons, the following units can be selected.
For Gas
:
%; %a.s; ppm (vol/vol)
O2
Temperature :
°C, °F
For liquids (DO)
O2
:
Temperature :
ppb; ppm (w/w) ; mg/l; µg/l
°C, °F
The units ppb; ppm; mg/l and µg/l are all oxygen values dissolved in liquid (DO).
4. You can return to the menu with “Enter”.
5. Then select the
icon to return to the “Configuration” menu
Info
This screen shows the software version and serial number present on the OGM.
Date
You can alter the date with the scroll buttons. By pressing “Enter” you can enter the date. By
pressing “Enter” again, you can enter the month and the year. Close with “Enter” to return to the
“Configuration” menu.
Time
You can alter the time with the scroll buttons. You can set the hours with “Enter”. By pressing
“Enter” again, you can enter the minutes and seconds. Close with “Enter” to return to the
“Configuration” menu.
OGM
Page 23 of 81
HAFFMANS OGM
Language
With the scroll buttons, you can choose Dutch, German, English, French and Spanish; select a
language by pressing “Enter”, you are then returned to the configuration menu.
Sensor settings
In the “Sensor Settings” menu, you can indicate for each sensor whether or not this is
connected, and whether these are to be used for liquids or gas. The sample time can also be
set for each sensor. Apart from entering the sample time, the sensors can also be set in auto
mode. If the auto mode is chosen, the sample time is chosen automatically. This shall start
standard with a sample time of 2 seconds. If the measured value deviates by more than 10%
after this time, the sample time remains 2 seconds. If the sample time deviates by less than 5%,
this will be doubled to 4, 8 15 etc. to a maximum of 60 seconds. So if in the next measurement:
Deviation < 5%, the sample time shall be, resp. 2 4 8 15 30 60 seconds.
Deviation > 10%, the sample time shall be, resp. 60 30 15 8 4 2 seconds.
The following lists the setting options:
Item
Setting options
Default setting
O2 sensor 1
OFF, ON
ON
O2 sensor 2
OFF, ON
OFF
Item
range
default
Unit
Sample time (sensor 1)
AUTO, 2 … 999
30
s
Sample time (sensor 2)
AUTO, 2 … 999
30
s
Gas application:
The O2 sensor selectively measures the partial oxygen pressure.
Gas application:
The oxygen value is expressed standard at atmospheric pressure.
The partial oxygen pressure is directly proportional to the gas pressure.
To attain the proper O2 result at atmospheric pressure, of a gas measured under pressure, the
measured O2 value has to be compensated.
Example:
Patm during calibration: 1013 mBar
O2 measured at 1.033 Bara (20 cm H2O) = 1,0 %
O2 at Patm= 0.981 %
Liquid application:
If the O2 sensor is used in liquid, this is fully pressure independent.
The settings can be set/altered as follows:
1. Select “Sensor Settings” in the “Configuration” menu and press “Enter” once.
The menu “Sensor Settings” appears.
Page 24 of 81
OGM
HAFFMANS OGM
2.
3.
4.
5.
Select the desired item in this menu with the cursor buttons and confirm with “Enter”.
The desired settings can then be chosen with the cursor buttons.
Use the “Enter” button to return to the menu “Sensor Settings”.
Then use the cursor buttons to select the
icon, and return to the “Configuration”
menu by pressing the “Enter” button.
Automatic switch-off on temperature.
This function is described as “auto stop temp”. The function of this parameter is to automatically
stop the O2 measurement on reaching this temperature and is typically used to prevent ageing
of the O2 coating during cleaning, without offering a remote signal. If the temperature again
drops below this set value with a hysteresis of 1 °C , the sensor shall automatically resume
measuring. This function applies to both sensors.
Item
Range
Default setting
Auto stop temp
0°C….90°C
50°C
Analogue outputs
The analogue outputs can be set as desired. A maximum of two analogue outputs can be set.
After selecting “Analogue Outputs”, the attendant setting of the unit automatically appears (after
confirmation). Depending on the units set under “Units”. The range is also adapted if the unit is
altered. The following lists the settings and the default settings.
Item
Setting options
Default setting
Configure output 1
O2 sensor1, O2 sensor2, Temp.sensor1, Temp.sensor2
O2 sensor 1
Configure output 2
O2 sensor1, O2 sensor2, Temp.sensor1, Temp.sensor2
Temp. sensor 1
Item
Setting options
Default setting
Levels Standby
Min, Max, Freeze
Freeze
Item
Range
Default
Unit
Minimum value ( if
O2 sensor)
0 …. 45000
0
Maximum value (if
O2 sensor)
0 …. 45000
2000
ppb;
ppm; mg/l; ug/l; %as; %
ppb;
ppm; mg/l; ug/l; %as; %
Minimum value (if
Temp.sensor)
- 10 … 150
-5
Maximum value (if
temperature sensor)
- 10 … 150
OGM
°C
°F
110
°C
°F
Page 25 of 81
HAFFMANS OGM
The settings can be set/altered as follows:
1. Select “Analogue Output” in the “Configuration” menu and press the “Enter” button once.
The menu “Analogue Outputs” appears.
2. In this menu, select the option “Limits” with the cursor buttons and confirm with “Enter”.
3. Then use the cursor buttons to select the setting concerned and alter these settings
where required.
4. Use the “Enter” button to return to the “Analogue Outputs” menu.
5. Use the cursor buttons to select the
icon and use the “Enter” button to return to
the “Configuration” menu.
mA Values output 1 and 2
The mA values of output 1 and 2 are controlled as follows:
Status
Off
Init
Standby
Measure
Calibrate
Autostop
O2 signaal
No signal
4mA
Adjustable (Min, Max, freeze)
Measurement value = mA
20mA
Adjustable (Min, Max, freeze)
Temperatuur signaal
No signal
4mA
Measurement value = mA
Measurement value = mA
Measurement value = mA
Measurement value = mA
Setting = “Min”: ‘The mA values of output 1 and 2 go to 4mA in status Standby or Autostop
Setting = “Max”: ‘The mA values of output 1 and 2 go to 20mA in status Standby or Autostop
Setting = “Freeze”: The mA values of output 1 and 2 keep the actual measurement value in
status Standby or Autostop
In case of an alarm output 1 and 2 will be controlled as shown in chapter 9.2 (Alarm messages)
Note!
Page 26 of 81
The settings for output 1 and 2 determine the measured values for
the Analog outputs. See I/O connector 5: Pin 7+8 in appendix C.
OGM
HAFFMANS OGM
Alarm settings
As soon as the electronics discovers an alarm, this is indicated with a flashing backlight on the
display. The lower part of the display shows which alarm was discovered. In the “Alarm” menu,
the alarm levels concerning the measured values can be set as desired. The settings apply to
sensor 1 and sensor 2. If only one sensor is connected, the other settings
will not be visible. The following list the settings options and the default settings.
Range
Item
Default
unit
O2 alarm
0 …. 45000
500
ppb;
ppm; mg/l;
ug/l; %as; %
O2 extra
0 …. 45000
1000
ppb;
ppm; mg/l;
ug/l; %as; %
Temperature low
- 10 … 150
20
°C
°F
Temperature high
- 10 … 150
20
°C
°F
The settings can be implemented as follows:
1. Select “Alarm Settings” in the “Configuration” menu and press the “Enter” button once.
The menu “Alarm Settings” appears.
2. Select the desired item in this menu with the cursor buttons and confirm with “Enter”.
3. Then select the setting concerned with the cursor buttons.
4. Use the “Enter” button to return to the “Alarm Settings” menu.
5. Use the cursor buttons to select the
icon and use the “Enter” button to return to
the “Configuration” menu.
Alarm outputs
The following four alarm outputs are available:
A1: General alarm
A2: O2 alarm
A3: O2 extra alarm sensor 1
A4: O2 extra alarm sensor 2
An O2 alarm can be set for each O2 measurement in two ways:
- high – high/high, if an O2 level should not be exceeded, where two levels can be set, e.g.
high as alarm and high/high to intervene in the process.
- low – high, if the O2 level has to be within a range, typically applied for yeast activation.
OGM
Page 27 of 81
HAFFMANS OGM
Conditions alarm type:
Sensor 1
O2 high > O2 extra
O2 high > O2 extra
Sensor 2
O2 high > O2 extra
O2 high > O2 extra
Type of alarm
High - high/high
low - high
Digital outputs
A1, A2, A3
A1, A2, A3
High - high/high
Low - High
A1, A2, A4
A1, A2, A4
General alarm (A1):
- All defects
- All error messages
- Temperature alarms of sensor 1 and/or sensor 2
- O2 alarms of sensor 1 and/or sensor 2
Digital output A1 (Normally Close contact (NC)) will only go to zero if the alarm is present during
two measurements. The alarm can be reset with ACK/RESET ALARM in the shortcut menu.
This maintains these digital outputs on high (and shall remain high as long as the alarm is on).
The switching conditions of alarm outputs A2, A3 and A4 can easily be found in the following
information.
Sensor
Alarm
A1
A2
A3
A4
1
Temp < > min/max
x
.
.
.
1
High-high/high:
O2 > O2 high
x
x
.
.
1
High-high/high:
O2 > O2 extra
x
x
x
.
1
Low-high:
O2 > O2 high
x
x
.
.
1
Low-high:
x
.
x
.
O2 > O2 extra
1
Sensor fault / defect
x
.
.
.
2
Temp < > min/max
x
.
.
.
2
High-high/high:
O2 > O2 high
x
x
.
.
2
High-high/high:
O2 > O2 extra
x
x
.
x
2
Low-high:
O2 > O2 high
x
x
.
.
2
Low-high:
O2 > O2 extra
x
.
.
X
x
.
.
.
2
Page 28 of 81
Sensor fault / defect
OGM
HAFFMANS OGM
High – high/high alarm (O2 extra > O2 high)
O2 extra
O2 high
Measured
value
t [min]
A3
extra alarm
output
A2
alarm output
A1
Alarm on
fail-save
2
2
Ack/reset
Alarm
OGM
2
2
Ack/reset
Alarm
Ack/reset
Alarm
Page 29 of 81
HAFFMANS OGM
Low – high alarm (O2 extra <O2 high)
O2 High
low
(O2 extra)
Measured
value
t [min]
A3
extra output
A2
alarm output
A1
Alarm on
fail-save
2
2
Ack/reset
Alarm
2
2
Ack/reset
Alarm
For a list of the physical connections of the I/O see appendix C.
All digital outputs have been executed as normally closed (NC).
The alarms go to zero if the alarm is present during two measurements.
Ack/reset Alarm:
The alarm can be reset with ACK/RESET ALARM in the shortcut menu. This maintains these
digital outputs on high (and shall remain high as long as the alarm is on). In other words, the
alarm relay shall fall out. Also, the backlight on the screen shall stop flashing. As long as the
alarm condition is present, this message shall remain on the alarm line on the screen.
This message shall disappear if the alarm condition is no longer present or if the measurement
is discontinued.
Page 30 of 81
OGM
HAFFMANS OGM
6.
Operation
The instructions in this chapter have to be performed before using the OGM for the first time.
This chapter also describes the instructions for daily use and operation.
6.1
Operation OGM
6.1.1 Main measuring screen
After the control box has been started up, the main measuring screen appears automatically
after 3 seconds. Depending on the number of connected sensors, a measuring screen can be
chosen:
Measuring screen sensor 1.
Measuring screen sensor 2.
Measuring screen sensor 1+2.
With the cursor buttons “sensor”, you can chose these screens. For further explanation, see
chapter 5.3.
6.1.2
Starting the measurement.
When the “Start” button is activated, the OGM shall start the measurement, and the red LED on
the start button shall light up. The measuring frequency during the measurement depends on
the sample time set during configuration, see chapter 5.5.2. “sensor settings”.
In the following, a survey of the start button function:
One sensor connected: Start button is to start/stop the measurement
Two sensors connected:
Measuring screen sensor 1: Start button is to start/stop measurement sensor 1
Measuring screen sensor 2: Start button is to start/stop measurement sensor 2
Measuring screen sensor 1+2: Start button is to start/stop measurement sensor 1+2
Note!
The red LED will light up if one or both sensor measurements are
conducted.
The last 400 measurements of each sensor are stored in the memory. If the memory is full, the
oldest measurement is overwritten by the latest.
OGM
Page 31 of 81
HAFFMANS OGM
6.1.3 Shortcut
With the - button, you can switch to a shortcut menu. With this menu, you can switch directly
to the measuring values, trend lines and sensor-specific values. See the following overview
below. With the cursor buttons, you can choose the screens you want. To confirm a selection ,
please press “Enter”.
Fig. 6.1: Survey shortcut menu.
6.1.4 Main menu
To alter various settings, please go to the main menu. To
access the main menu, press the ENTER button for 3
seconds. The screen as seen in fig. 6.2 appears.
Fig. 6.2: Main menu
Page 32 of 81
OGM
HAFFMANS OGM
6.2
Local and Remote control
Before using the OGM, all desired parameters have to be set locally on the control unit. You can
then choose whether to operate the sensor locally or remote. If you choose remote, the control
unit can no longer be operated locally (start/stop). For remote control, you can choose between:
- RS 232 (protocol delivered on request)
- Profibus
6.3
Logbook
You can choose the logbook screen from the main menu. This screen shows various values,
statistics and charts. For a comprehensive overview of the screens in the logbook, see
appendix B “Software structure”. A short overwiew of these screens is given below.
Measuring values
This menu shows the measuring values of sensors 1 or 2 separately. No more than 400
measurements are stored in the memory. For each measuring value, the O2 value,
temperature, date and time is stored. If the memory is full, the oldest measurement is replaced
by the latest.
Trending
A chart can be shown for the measured oxygen values as well as for the temperature. The
values shown are always the measurements for the last 30 minutes.
Depends on the unit chosen in “Units”.
O2 (0,5 % / DIV)
Depends on the max. value chosen in analogue outputs.
5
extra: O2 extra alarm value
extra
max. O2 extra alarm value
max.
0
OGM
0
Time (1,0 min. / DIV)
30
Page 33 of 81
HAFFMANS OGM
Operational hours
This menu shows the hours when the sensor is operational. This distinguishes between
measuring hours and standby hours.
Extremes
This menu stores the extreme values of the oxygen and the temperatures.
Calibration
The last three calibration data of the O2 sensor are stored. The last calibration data of the
sensor are stored in the memory of the control unit. If a “Restore” is input in the sensor, the last
calibration data of the sensor are restored. This option is no longer functional (i.e. the calibration
data are incorrect) if a sensor or control unit is exchanged between two calibrations.
Alarms
In this menu, no more than 50 alarms are shown, with the date and time. The user cannot adapt
this menu. If the memory is full, the oldest alarm is replaced by the latest.
Reset
All stored measurements can be deleted, using reset.
Page 34 of 81
OGM
HAFFMANS OGM
7.
Calibration
7.1
Calibration of the temperature sensor
The temperature sensor is an accurate NTC with precise tolerances, and therefore does not
need to be calibrated.
7.2
Calibration of the O2 sensor(s)
Because the oxygen sensor is only subject to very minor ageing due to light exposures, regular
calibration is not necessary. Research has shown that the LHO coating, after 1 million times
light exposure (i.e. 1 million O2 measurements), the O2 deviation from the zero point is only 5
ppb, and no deviation is found with higher O2levels.
There is a light pulse counter integrated in the software, which counts the number of light
exposures. If there have been 1 million light exposures, the warning “calibrate or replace O2
sensor” and the sensor has to be recalibrated or replaced. As a reminder, the display
continuously shows a warning symbol.
O2
After succesful calibration, the light pulse counter is reset and the warning symbol disappears.
The complete light pulse counter is not reset. If a complete new coating holder is installed, this
is reset (after selection).
In the following, three primary options to calibrate the oxygen sensor are addressed:
1. Calibration-free
This method is recommended if no calibration gases are available. The oxygen sensor is
replaced with a pre-calibrated oxygen sensor, delivered with certificate.
2. One-point calibration
LHO: 0-point calibration accurately in the lower oxygen level and less accurately in
higher oxygen level.
LHG: 0-point calibration accurately in the lower oxygen level and less accurately in
higher oxygen levels.
WLO: air-calibration accurately in the higher oxygen level and less accurately in lower
oxygen levels.
3. Calibration with two gases.
This method is recommended to guarantee the specified accuracy for the entire oxygen
measuring scope.
7.2.1 Calibration frequency
Recalibration is recommended after 1 million exposures. When using the standard sample time
(1 measurement / 30 seconds), the sensor has to be calibrated once a year, assuming the
sensor has then been operational 24/7. If a quicker sample time is required, the sensor will
reach 1 million exposures soon, and calibration is then necessary.
OGM
Page 35 of 81
HAFFMANS OGM
7.2.2 Calibration-free
The “Calibration free” method is applicable when changing the O2 sensor and the new O2
sensor has a calibration certificate. For the “calibration free” calibration, the calibration data for
the zero point and the high calibration point can be manually entered into the software.
Procedure
1. Press the “Enter” button in the main measuring menu longer than 3 seconds.
The main menu appears.
2. Now use the cursor buttons to select “Calibration” and confirm with “Enter”. (Please note:
the calibration menu is a level B supervisor level!)
3. Then select the sensor you want to calibrate and
confirm with “Enter”.
(“Latest calibrations” shows when the last
calibration was executed).
4. Then select “Manual Calibration” and
finish with “Enter”.
(The “Light pulse counter” shows how many light
pulses the coating had between two
intervals. The “Total light pulse counter” shows how
many light pulses the coating had from
when it was first operational. The other data are the
current measured values of the oxygen sensor).
5. Now select consecutively the atmospheric pressure, the
phase angles, the temperatures and the oxygen
concentrations, recorded on the certificate supplied by
Haffmans BV. Enter the numbers, using the cursor
buttons, use the “Enter” button to go to the next number.
After the last number, press the “Enter” button to go to
the next item on the menu.
6. After entering the calibration data for replaced coating, select “Yes” on the line “Replace
coating”.
7. Then select “store calibration” and the calibration values are stored in the O2 sensor.
With the
icon you will return to the previous menu.
Please note!
Page 36 of 81
If you return to the previous menu without selecting “store
calibration”, the old calibration data are maintained.
OGM
HAFFMANS OGM
7.2.3 One-point calibration
LHO & LHG –sensor
For most calibrations, the low oxygen level is usually sufficient to recalibrate the zero point.
To do so, follow the instructions in chapter 7.2.4 “Calibration with two gases”. It is not necessary
to enter the atmospheric air pressure and the high calibration point again. Therefore, skip points
7, 9 and 10.
WLO –sensor
For most applications in high oxygen measuring level, it is usually sufficient to recalibrate the
high calibration point in the air.
To do so, follow the instructions in chapter 7.2.4 “Calibration with two gases”. It is not necessary
to enter the low calibration point again. Therefore, skip points 8 and 10.
7.2.4 Calibration with two gases
Additional requirements:
Barometer with accuracy of one mbar.
For the WLO sensor, a relative air humidity measurement with
an accuracy of 1% RV is necessary when calibrating
with a high oxygen level calibration point with liquid gas.
When calibrating, the zero point and a high calibration point can be calibrated with a certified
oxygen concentration. The calibration gases have to be dry gases. The following lists the gases
for the different sensor types:
Sensor type
LHO
0-point
N2 class 5.0, CO2 class 5.0
LHG
N2 class 5.0, CO2 class <0.5 ppm O2 !
CO2 class 5.0 O2 class < 0.5 ppm O2 !
Or use class 6.0 N2 or CO2
N2 class 5,0, CO2 class 5,0
WLO
High point
Gas mixture O2 concentration
1 - 4%
Accuracy 1%
Gas mixture O2 concentration 100
– 200 ppm
Accuracy 1%
Air (with 20,95 % O2)
Table 7.1: list of gases for calibration
The OGM has to be dry when calibrating with gas. Flush the OGM for approx. 15 min. with N2 or
CO2 or oil-free compressed air. During calibration, the unit % or %a.s. has to be used for the
LHO and WLO coating. For the LHG coating, always use the unit PPM (gas)! Because the
oxygen sensor measures the partial oxygen pressure, the oxygen level measured in gas during
calibration depends on the current air pressure.
By manually entering the current air pressure in the parameter “P-atm” in the calibration menu,
when showing the oxygen level in the gas unit, %O2 and % a.s. are taken into account.
Note!
OGM
For an accurate calibration, the calibration gas and the OGM have
to acclimatize for 4 hours in the same room.
Page 37 of 81
HAFFMANS OGM
Procedure
1. Place the OGM in the calibration beaker intended for that purpose.
2. Press the “Enter” button in the main measuring menu longer than 3 seconds.
The main menu appears.
3. Now select “Calibration” with the cursor buttons and close with “Enter”. (Please note: the
calibration menu is a level B supervisor level!)
4. Then select the sensor you want to calibrate and confirm
with “Enter”.
(“Latest calibrations” shows when the last
calibration was executed)
5. Then select “O2 auto” and confirm with “Enter”.
(The “Light pulse counter” shows how many light pulses
the coating had between two intervals. The “Total light
pulse counter” shows how many light pulses the coating
had from when it was first operational. The other data
are the current measured values of the oxygen sensor).
6. The software automatically recognizes which coating
is present on the sensor.
(LHO, LHG or WLO)
7. Then enter the current air pressure.
To do so, use a barometer that
is accurate to one millibar.
Page 38 of 81
OGM
HAFFMANS OGM
8. Then flush out the calibration beaker
with the right type of gas, see table 7.1.Then let
the gas flow for 15 minutes with a
maximum speed of 30 l/h. Now, press “Enter”,
so the cursor is at value 0.00%.
Press enter again, to store the calibration
values of the zero point. The calibration
value is now stored in the memory.
If no stable value is reached,
the message: ”O2 unstable values” appears.
9. Then calibrate the high calibration point.
Connect the right type of calibration gas to the
calibration beaker and let this flow for 15 minutes,
with a maximum flow of 30l/h. Press “Enter” again
and enter the oxygen level of the calibration gas,
and again close with “Enter”.
The calibration value is now stored
in the memory. If no stable value is reached,
the message :”O2 unstable values” appears.
For the WLO-sensor, the relative humidity has to be taken into account. This value has
to entered in “RV-act”. The outside air or dry compressed air contains 20.95% oxygen.
10. Then indicate whether or not you have replaced the
coating. If you enter “Yes”, the “Total light pulse counter”
is set to zero.
11. If you have a new coating holder with new coating
constants, select the menu “Constants”. The coating
holder constants screen appears. Enter the new
constants and then return to the calibration menu.
OGM
Page 39 of 81
HAFFMANS OGM
12. Now select “Store calibration” and the data
are stored permanently in the sensor
Then use the
icon to return to
the calibration menu.
Note!
If you return to the previous menu without selecting “store
calibration”, the old calibration data are maintained.
Checking calibration data.
If you want to check the OGM for its calibration point with
calibration gases, you can also use the calibration menu.
Connect the gases to the calibration beaker and enter
the right Patm in the calibration menu.
The right oxygen value is then calculated in “Actual O2 value”.
Please note!
Page 40 of 81
When checking, do not use “store calibration”
OGM
HAFFMANS OGM
7.3
Calibrating analogue outputs
The analogue outputs of the OGM control unit can be calibrated.
These values then match the values entered in Analogue Outputs.
The analogue outputs are calibrated standard at 4mA and 20mA.
Procedure
1. Press the “Enter” button in the main measuring menu longer than 3 seconds.
The main menu appears.
2. Now select “Calibration” with the cursor buttons and close with “Enter”. (Please note: the
calibration menu is a level B level!)
3. Then select “Analogue outputs” with the cursor buttons and close with “Enter”.
4. Now select “Calibration analogue outputs” with the
cursor buttons and close with “Enter”.
5. The analogue outputs can now be calibrated in this
screen for the right mA-controls. Please connect an
mA-meter (or multimeter) to the output concerned
to measure the current in the output.
See appendix C for connections of the analogue outputs.
6. By selecting the “MIN. mA-value”, the
connected analogue output is controlled. With the
cursor buttons you can increase or decrease
the value set. Increasing /decreasing the count value
shall run concurrently with increasing/decreasing the
current. This calibrates the output.
7. By selecting the “Max. mA-value”, the
connected analogue output is controlled. With the
cursor buttons you can increase or decrease
the value set. Increasing /decreasing the count value
shall run concurrently with increasing/decreasing the
current. increasing/decreasing the current. This
calibrates the output.
8. After calibrating the outputs, select “Store calibration” to store the data in the OGM
memory. Then use the
icon to return to the previous menu.
Note!
OGM
If you return to the previous menu without selecting “store
calibration”, the old calibration data are maintained.
Page 41 of 81
HAFFMANS OGM
Page 42 of 81
OGM
HAFFMANS OGM
8.
Cleaning and maintenance
8.1
Daily maintenance
8.1.1 Inspection
•
•
•
8.1.2
Ensure that the instrument is free of dirt and dust.
Check that plugs and cables are properly connected.
Blind off connectors not in use.
Cleaning
•
•
The LHG version can not be cleaned in the pipes with regular CIP.
Only clean with a cloth, and water if necessary.
LHO + WLO versions
Regular CIP of the process pipes with appropriate cleaning agents.
Agent - Alkali cleaning based on NaOH of KOH
(≤95°C, ≤5% NaOH or KOH)
- Acidic cleaning based on HNO3 of H3PO4
(≤60°C, ≤3% HNO3 or H3PO4)
•
Current
0 - 10 m/s
Clean the sensor or the control unit on the outside just with water (possibly with a
detergent).
Warning
Never direct a powerful jet of water at the control panel
or at other instruments.
Warning
Never use tools to clean the coating. Touching the
coating can cause serious damage to the instrument
OGM
Page 43 of 81
HAFFMANS OGM
Page 44 of 81
OGM
HAFFMANS OGM
9.
Faults and repair
9.1
List of faults
Problem
Cause
Solution
Nothing is displayed on the
screen when a button is
pressed.
The power is not
connected.
The power is in the safemode (too hot).
Connect the power.
The O2 value seems to be
wrong.
Defective temperature
sensor.
The message “O2 sensor is
not responding” appears on
the measuring screen
Oxygen sensor not
connected.
Oxygen sensor defective.
9.2
Alarm messages
ID
Error message
Cause
Wait 10 minutes and connect the power
again.
Check whether a realistic temperature
value is indicated.
Check O2 sensor with calibration gases
and possibly recalibrate.
Connect oxygen sensor.
Contact Haffmans BV.
Solution
mA Output
20mA
01
S1 – Temp sensor not
found
NTC defective
Install new NTC, contact
Haffmans BV
02
S1 – Temp sensor
short circuit
NTC short-circuit
Install new NTC, contact
Haffmans BV
No sensor connected,
Sensor defective, Break
in communication cable,
Thermal disconnection
sensor (T>65°C)
No sensor connected
20mA
Check connection
Reset OGM
Let the sensor cool down
03
S1 – O2 comm error
04
S1 – Not found
05
S1 – Temperature high
Temp. Value > set “temp
high” value
06
S1 – Temperature low
Temp. Value < set “temp
low” value
07
S1 – O2 high alarm
O2 value > set “O2 high"
alarm
Check the process, alter
the alarm “O2 high”
08
S1 – O2 extra alarm
O2 value > set “O2 high”
alarm
Check the process, alter
the alarm “O2 extra”
09
S1 – Calibration
expired
O2
The coating reached 1
million light pulses.
The sensor has to be
recalibrated.
OGM
20mA
Connect sensor
Check the process
Alter the alarm “temp
high” level
Check the process
Alter the alarm “temp low”
level
20mA
Actual
measurement
value
Actual
measurement
value
Actual
measurement
value
Actual
measurement
value
Actual
measurement
value
Page 45 of 81
HAFFMANS OGM
ID
Error message
Cause
Solution
10
S2 – Temp sensor not
found
NTC defective
Install new NTC, contact
Haffmans BV
11
S2 – Temp sensor
short circuit
NTC short-circuit
Install new NTC, contact
Haffmans BV
20mA
Check connection
Reset OGM
Let the sensor cool down
20mA
No sensor connected,
Sensor defective, Break
in communication cable,
Thermal disconnection
sensor (T>65°C)
No sensor connected
mA Output
20mA
12
S2 – O2 comm error
13
S2 – Not found
14
S2 – Temperature high
Temp. Value > set “temp
high” value
15
S2 – Temperature low
Temp. Value < set “temp
low” value
16
S2 – O2 high alarm
O2 value > set “O2 high"
alarm
Check the process, alter
the alarm “O2 high”
17
S2 – O2 extra alarm
O2 value > set “O2 high”
alarm
Check the process, alter
the alarm “O2 extra”
18
S2 – Calibration
expired
The coating reached 1
million light pulses.
The sensor has to be
recalibrated.
19
S1 – Init error
20
S2 – Init error
Not all calibration data
have been entered
correctly.
Not all calibration data
have been entered
correctly.
Connect the sensor to the
control unit and reset the
control unit.
Connect the sensor to the
control unit and reset the
control unit.
23
I/O communication
failed
Electronics defective.
Contact Haffmans BV
20mA
Compare the measured
temperature with the
temp. set at “auto stop
temp.” and, if required,
increase this.
Adjustable
Compare the measured
temperature with the
temp. set at “auto stop
temp.” and, if required,
increase this.
Adjustable
24
25
O2
S1 – Auto Stop Active
S2 – Auto Stop Active
Page 46 of 81
One temp. > then the
temperature set at
parameter “auto stop
temp” is measured,
switching on the auto
stop function of the O2
measurement.
One temp. > then the
temperature set at
parameter “auto stop
temp” is measured,
switching on the auto
stop function of the O2
measurement.
Connect sensor
Check the process
Alter the alarm “temp
high” level
Check the process
Alter the alarm “temp low”
level
20mA
Actual
measurement
value
Actual
measurement
value
Actual
measurement
value
Actual
measurement
value
Actual
measurement
value
20mA
20mA
OGM
HAFFMANS OGM
9.3
Repairs and maintenance
When assembling and disassembling, make sure that there is
no power on the instrument,
In such a manner that there is no risk of electric shocks.
Warning
To prevent damage, first read the instructions as described in
this manual
Replacing the coating holder:
If the coating is defective or damaged, a new coating holder of the same type has to be placed
on the sensor.
Instructions for replacing the coating holder:
1. Take the sensor from the pipe and place it on a flat straight surface so it cannot roll from
the workplace onto the ground.
2. Screw the coating holder from the
measuring head with a wrench/ring
spanner size 14. See fig. 9.1. You can only
use a spanner for the WLO version.
The 25mm coating holders have a wrench
size 10.
Fig. 9.1: Unscrewing the coating holder
Light
3. Remove glue and dirt residue from the
screw thread around the light conductor
Please note! Do not damage the light
conductor See fig. 9.2.
Screw thread
Fig. 9.2: Light conductor
OGM
Page 47 of 81
HAFFMANS OGM
4. Then take the new coating holder and
assemble the large O-ring around the flange
See fig. 9.3.
Fig. 9.3: New coating holder
5. Now apply a drop of Loctite 572 on the inside
of the screw thread of the coating holder. to
prevent the coating holder from possibly
shaking loose. Prevent excess glue from
blocking the path of the light.
Add Loctite 572.
Fig. 9.4: Apply the coating holder
screw with a drop of loctite
6. Then screw the coating holder back on the sensor and calibrate it.
Warning
Page 48 of 81
Never unscrew the temperature sensor. If you unscrew the
temperature sensor, you forfeit your warranty right.
OGM
HAFFMANS OGM
10.
Deactivating, storage and transport
10.1
Deactivating
If the OGM is no longer used, the components must be sent to an appropriate waste disposal
organization or the entire unit can be returned to Haffmans BV. The materials from which the main
components are made are given below:
-
10.2
Control unit
Sensor
: ABS casing, various prints.
: Stainless steel, oxygen print, POM
Storage
The instrument must be stored in a dry location and be protected against mechanical and/or
chemical damage. The calibration beaker has to be installed to protect the coating.
10.3
Transport
When the OGM is sent, the display of the control unit has to be protected against direct
mechanical shocks.
The O2 sensor also has to be provided with the calibration beaker, to protect the O2 coating.
We advise keeping Haffmans’ packaging materials for reuse, if the OGM has to be sent
someplace.
OGM
Page 49 of 81
HAFFMANS OGM
Page 50 of 81
OGM
HAFFMANS OGM
Appendix A: Formulas
Conversion O2 units
The OGM can be set for various oxygen units, where the following conversions can be
conducted in the instrument:
%O2 = % air saturation ⋅
20,95
100
With
Tm :
Temperature in ° C
patm : Atmospheric air pressure during the oxygen calibration in bar
(
)
cT = 48,998 - 1,335 ⋅ Tm + 0,02755 ⋅ Tm2 - 0,000322 ⋅ Tm3 + 0,000001598 ⋅ Tm4 temperature
correction
mg
%air saturation 0,2095
32
= p atm ⋅
⋅
⋅ cT ⋅
l
1013
100
22,414
mg
l
µg mg
=
⋅ 1000 = ppb
l
l
ppm =
OGM
Page 51 of 81
HAFFMANS OGM
Page 52 of 81
OGM
HAFFMANS OGM
Appendix B: Software structure
Starting screen
Automatic
After 3 sec.
Button ↓
Button ↓
Button ↓
Sensor 2
Sensor 1
Sensor 1 & 2
Measuring
screen
↓
button
→ Button
Main menu
Measuring values
Sensor 1
O2 trend chart
Sensor 1
Details values
Sensor 1
Shortcut
Measuring values
Sensor 2
O2 trend chart
Sensor 2
Details values
Sensor 2
OGM
Page 53 of 81
HAFFMANS OGM
A = Operator level
B = Supervisor level
User
A
Units
B
Info
Configuration
A
A
Sensor settings
Main menu
B
Analogue outputs
B
Limits
B
Calibration
analogue outputs
B
Alarm settings
B
Auto Calibration
sensor 1
B
Calibrations
B
Calibration
sensor 1
B
Regular
calibration
sensor 1
Calibration
sensor 2
B
Auto Calibration
sensor 2
B
Regular
calibration
sensor 2
Page 54 of 81
OGM
HAFFMANS OGM
Measuring values
sensor 1
A
Measuring values
A
Measuring values
sensor 2
A
Trending
A
O2 chart s1
Display
A
Company
A
Temp chart s1
Display
A
O2 chart s2
Display
A
Extremes
A
Temp chart s2
Display
A
Calibrations
A
History1 Sensor1
A
Logbook
A
History2 Sensor1
A
Alarms
A
Reset
B
History3 Sensor1
A
History1 Sensor2
A
History2 Sensor2
A
History3 Sensor2
A
OGM
Page 55 of 81
HAFFMANS OGM
Page 56 of 81
OGM
HAFFMANS OGM
Appendix C: Connections
Connector:
M/S Power supply
A
RS-communication / Profibus communication
B
Sensor communication 1
C
Sensor communication 2
D
I/O communication
Tabel 1
M
1.
2.
3.
OGM
Mains 230Vac
Mains supply
Mains supply
Protective earth
3-way
L1
L2
PE
Female (code 1)
Brown
Blue
Yellow-green
M
Page 57 of 81
HAFFMANS OGM
S
1.
2.
3.
Supply 24Vdc
External supply
External supply
Protective earth
3-way
+24V
GND
PE
Male (code 2)
Red
Black
Yellow-green
A
1.
2.
3.
4.
5.
6.
Communication
Common
Supply, 24Vdc
COM-1, RS232
COM-1, RS232
COM-1, RS232
Supply, 12Vdc
6-way
GND
+24V
Txd-1
Rxd-1
Gnd-1
+12V
Male (code 1)
White
Brown
Green
Yellow
Grey
Pink
B
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
0.
Sensor 1
Common
Supply, 24Vdc
COM-2, RS232
COM-2, RS232
COM-2, RS232
Temp. input-1*
Temp. ground
Supply, 12Vdc
COM-2, RS485
COM-2, RS485
Shielding
10-way
GND
+24V
Rxd-2
Txd-2
Gnd-2
NTC-1
NTC-com
+12V
B (D+)
A (D-)
Ground
Male (code 2)
White
Brown
Green
Yellow
Grey
Pink
Blue
Red
Black
Violet
Casing
S
A
B
* In use for OGM only (not available at OptHaze-i)
Page 58 of 81
OGM
HAFFMANS OGM
C
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
0.
Sensor 2
Common
Supply, 24Vdc
COM-3, RS232
COM-3, RS232
COM-3, RS232
Temp. input-2*
Temp. ground
Supply, 12Vdc
COM-2, RS485
COM-2, RS485
Shielding
10-way
GND
+24V
Rxd-3
Txd-3
Gnd-3
NTC-2
NTC-com
+12V
B (D+)
A (D-)
Ground
Male (code 2)
White
Brown
Green
Yellow
Grey
Pink
Blue
Red
Black
Violet
Casing
C
* In use for OGM only (not available at OptHaze-i)
D
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
0.
PLC I/O
Common - PLC *
DON’T USE **
Sensor 1 on/off
Sensor 2 on/off
Set product ID1•
Set product ID2•
Spare input
Spare input
GND for inputs
An. output-1 +
An. output-1 –
An. output-2 +
An. output-2 –
Alarm output-1
Alarm output-2
Alarm output-3
Alarm output-4
Input value +
Input value –
Shielding
19-way
GND
+24V
Input-24V
Input-24V
Input-24V
Input-24V
An-input
An-input
Gnd-IO
Out-20mA
Out-gnd
Out-20mA
Out-gnd
Outp-24V
Outp-24V
Outp-24V
Outp-24V
Inp-20mA
Inp-20mA
Ground
Male (code 2)
White
Brown
Green
Yellow
Grey
Pink
Blue
Red
Black
Violet
Grey-pink
Red-blue
White-green
Brown-green
White-yellow
Brown-yellow
White-grey
Brown-grey
White-pink
Casing
D
* Use as GND (reference) to PLC and/or for alarm relays
**
DON’T USE to/for supply (reference for inputs only)
•
Use for OptHaze-i only (not available at OGM)
OGM
Page 59 of 81
HAFFMANS OGM
Example connection I/O:
Page 60 of 81
OGM
HAFFMANS OGM
OGM
Page 61 of 81
HAFFMANS OGM
Page 62 of 81
OGM
HAFFMANS OGM
Appendix D: RS-protocol
Setting
Baudrate
Databit
Parity
Stopbit
Hardware control
57600
8
No
1
No
Note: Select used com-port in the Haffmans programm
OGM
Page 63 of 81
HAFFMANS OGM
Page 64 of 81
OGM
HAFFMANS OGM
Appendix E: Instruction for assembly panel
There are three options to install the control box.
1. Wall mounting
The control unit can be mounted against a flat internal or external wall using the mounting
bracket, on a pipe using the optional pipe mounting bracket The maximum distance between
the control unit and the sensor is 15 meters.
Wall mounting (4x)
Fig. 5.1: Wall mounting
2. Panel Assembly
To assemble the control unit in a switch box, use the panel assembly set. This set is optional
available and contains the following parts:
-
OGM
Switchbox bracket
Seal
Cylindrical screw M5X40 (2 pieces)
Cutting template
Page 65 of 81
HAFFMANS OGM
A hole has to be cut in the panel door. Use the included cutting template to define the
dimensions of the hole.
Stick the seal at the backside of the control unit. The control unit can now be mounted in the
panel door. Use the switchbox bracket to clamp the control unit to the inner side of the panel
door. To fit the door and the control unit together, use the cylindrical screws to strain the
switchbox bracket. The result will look like fig. E.1.
Fig. E.1: Control unit panel assembly
3. Pipe mounting
Using the pipe clamp set the control box can be mounted on a vertical or horizontal pipe. This set is
available as an option and includes the following components:
-
Pipe clamp with hinge
Clamping strip 2x
Cylindrical screw with nut (4 pieces)
The wall mounting support which is included as standard must be screwed to the pipe clamp using
the four cylindrical screws with nuts. This can be done in two ways depending on the choice of
mounting on a horizontal or vertical pipe. The pipe clamp can be mounted on the pipe using 2
clamping strips.
The hinge mounted on the pipe clamp is used to tilt the control box.
Page 66 of 81
OGM
HAFFMANS OGM
Mounting on vertical pipe
hinge
Fig. E.3: Control box vertical pipe mounting (with hinge)
Mounting on horizontal pipe
4x
cylindrical
screw
Fig. E.4: control box horizontal pipe mounting (without hinge)
OGM
Page 67 of 81
HAFFMANS OGM
Page 68 of 81
OGM
HAFFMANS OGM
Appendix F: Ultrapure Sensor including high pressure reducing set
(Optional)
F.1
Drawing
Pos.
Description
01
Welding socket 12mm
02
Ball Valve
03
Double nipple
04
Pressure reducing valve
05
Coupling
06
Steel hose
07
Needle Valve
08
Flowcel 25mm
09
O2 sensor
10
Cable for O2 sensor
11
Power supply cable
12
OGM Control panel
OGM
Page 69 of 81
HAFFMANS OGM
13
Profibus cable (option)
F.2
Mounting instructions
1.
2.
3.
4.
Weld the welding socket (pos. 1) on the pipe
Turn the ball valve and the pressure reducing valve (pos 02, 03, 04 en 05) in the
welding socket (pos 01), Attention: use Teflon tape.
Turn the hose (pos. 06) in the pressure reducing valve (pos 04 and 05) and on the
needle valve (pos 07). Attention: always use two wrenches for fastening and
loosening: wrench 12 on the turning part and an adjustable wrench on the housing.
Adjust the flow, with the reducing and needle valve, at a maximum of 30 l/h
Page 70 of 81
OGM
HAFFMANS OGM
Appendix G: Customer Service
Warning
Repairs to the instrument’s electronics may only be performed
by a Haffmans BV electrician.
The fault list includes instructions for solving small problems. If you are not able to solve a
problem, contact Haffmans BV, your local representative or your agent. Always state the name
of the instrument, the article number, the serial number and the Haffmans BV order or invoice
number.
If you have any questions concerning the control or operation of your OGM, please contact
Haffmans BV, your local representative or your agent.
HAFFMANS BV
Marinus Dammeweg 30
5928 PW Venlo The Netherlands
PO Box 3150
NL-5902 RD Venlo The Netherlands
T
F
E
I
OGM
+31 (0)77-3232300
+31 (0)77-3232323
[email protected]
www.haffmans.nl
Page 71 of 81
HAFFMANS OGM
Page 72 of 81
OGM
HAFFMANS OGM
Appendix H: Warranty conditions
The Haffmans BV warranty has precedence over all rights that the purchaser has towards his
supplier based on a sales contract between the purchaser and the supplier or based on the law.
Haffmans BV guarantees that, under normal use, the instrument requires no maintenance and
is free of material and manufacturing faults. This warranty is applicable for a period of 12
months from the date of the invoice.
Necessary recalibration and/or standard maintenance is not covered by this warranty. The
warranty is not applicable to fuses or an instrument or component which has been incorrectly
used, which has been adapted or which has been exposed to abnormal working conditions and
abnormal treatment.
If a fault is observed during the warranty period, then this must be reported in writing. The
instrument, together with the written description of the problem, must be returned to Haffmans
BV or to a Haffmans BV representative or agent.
The faulty components will be repaired or replaced free of charge provided that the fault has not
been caused by:
•
•
•
•
•
Using the instrument in a way that deviates from the instructions given in the instruction
manual.
A modification to the instrument.
Overdue maintenance.
Repairs by a third party that has not received authorization from Haffmans BV.
Abnormal work conditions or environmental conditions.
The purchaser will be charged for the labour costs for disassembly and assembly relating to the
repair, as well as for the transport costs for returning the instrument to the purchaser.
Haffmans BV accepts no liability for possible transport damage.
OGM
Page 73 of 81
HAFFMANS OGM
Page 74 of 81
OGM
HAFFMANS OGM
Appendix I: EC Declaration of conformity
OGM
Page 75 of 81
HAFFMANS OGM
Page 76 of 81
OGM
HAFFMANS OGM
Appendix J: Spare parts
If you order spare parts, we kindly request you to include all the information given below.
•
•
•
•
Serial number of the OGM
Position number
Description.
Article number.
Pos.
Description
Art. No.
O-ring 60x3 for varivent
109.065
O-ring 20x2.5 for 25mm sensor
112.157
Spare parts
Accessories & Repair parts
OGM
Coating holder set Varivent LHO
112.021
Coating holder set Varivent WLO
112.122
Coating holder set 25mm LHO
112.135
Coating holder set 25mm WLO
112.136
Coating holder set 25mm LHG
112.518
Control unit OGM wide range (100 – 240Vac) version 2
112.354
Control unit OGM wide range (100 – 240Vac) with Profibus version 2
112.356
Control unit OGM 24V version 2
112.355
Control unit OGM 24V with Profibus version 2
112.357
Mounting bracket assembly Controller unit AUCI/OGM
119.005
Mounting set for pipe AUCI
119.006
O2 sensor OGM/Vari/LHO IP67
112.024
O2 sensor OGM/25mm/LHO IP67
112.032
O2 sensor OGM/Vari/WLO IP67
112.026
O2 sensor OGM/25mm/WLO IP67
112.034
O2 sensor OGM/Vari/LHO IP68
112.305
O2 sensor OGM/25mm/LHO IP68
112.307
O2 sensor OGM/Vari/WLO IP68
112.306
O2 sensor OGM/25mm/WLO IP68
112.308
Calibration beaker set varivent, 2 spare o-rings and varivent clamp
112.010
Calibration beaker set 25mm with 2 spare o-rings
112.031
Page 77 of 81
HAFFMANS OGM
Page 78 of 81
Connector for wide range power supply 3p female
119.121
Connector for 24V DC power supply 3p male
119.122
Power supply cable wide range (100-240Vac) 5m, IP 67
119.132
Power supply cable wide range (100-240Vac) 15m, IP 67
119.133
Power supply cable wide range (100-240Vac) 40m, IP 67
119.134
Power supply cable 24V dc 5m, IP 67
119.136
Power supply cable 24V dc 15m, IP 67
119.137
Power supply cable 24V dc 40m, IP 67
119.138
Sensor communication cable 5m, IP 67 2x ODU 10-way
112.332
Sensor communication cable 10m, IP 67 2x ODU 10-way
112.333
Sensor communication cable 15m, IP 67 2x ODU 10-way
112.334
Sensor communication cable 5m with swivel, IP 67 2x ODU 10-way
112.335
Sensor communication cable 10m with swivel, IP 67 2x ODU 10-way
112.336
Sensor communication cable 15m with swivel, IP 67 2x ODU 10-way
112.337
Sensor communication cable 5m, IP 67 ODU 10-way – BINDER 8-way
112.901
Sensor communication cable 10m, IP 67 ODU 10-way – BINDER 8-way
112.902
Sensor communication cable 15m, IP 67 ODU 10-way – BINDER 8-way
112.903
Sensor comm. cable 5m with swivel, IP 67 ODU 10w - BINDER 8w
112.904
Sensor comm. cable 10m with swivel, IP 67 ODU 10w - BINDER 8w
112.905
Sensor comm.15m with swivel, IP 67 ODU 10w - BINDER 8w
112.906
Sensor communication cable 5m, IP 68 ODU 10-way – BINDER 8-way
112.911
Sensor communication cable 10m, IP 68 ODU 10-way – BINDER 8-way
112.912
Sensor communication cable 15m, IP 68 ODU 10-way – BINDER 8-way
112.913
Sensor comm. cable 5m with swivel, IP 68 ODU 10w - BINDER 8w
112.914
Sensor comm. cable 10m with swivel, IP 68 ODU 10w - BINDER 8w
112.915
Sensor comm.15m with swivel, IP 68 ODU 10w - BINDER 8w
112.916
Convertor BINDER 8p male – ODU 10p male
112.371
Convertor BINDER 8p female – ODU 10p male
112.374
Loose connector I/O 19p
109.130
I/O cable 5m, IP 67
109.141
I/O cable 15m, IP 67
109.142
I/O cable 40m, IP 67
109.143
Communication cable RS232 5m
119.163
OGM
HAFFMANS OGM
OGM
Communication cable RS232 15m
119.164
Profibusset OGM
112.361
CD-Rom, PC software for reading data, incl. RS-232 cable
By request
25mm/Varivent adapter, with cover pin set
112.231
25mm/Orbisphere adapter, with cover pin set
112.233
25mm/DN 25 Varivent adapter with cover pin set
112.234
Cover pin set
112.209
LHG set 25mm measuring unit
112.509
Page 79 of 81
HAFFMANS OGM
Page 80 of 81
OGM
HAFFMANS OGM
Appendix K: PROFIBUS
Information on connecting the controlbox to a Profibus system can be found in the manual of
the Profibus add on.
OGM
Page 81 of 81
Скачать